Phantom Study: Area and Clarity of Myocardial Wall Non-per Fused Defect Acquired using SPECT
Keywords:
Clarity. Defect area. Myocardial wall. Positioning. SPECTAbstract
Physician’s interpretation worked continuously to improve nuclear image to be of perfect quality. Single photon emission computed tomography (SPECT) is one of the most important methods to assess the diagnosis of diseases. Myocardial perfusion imaging (MPI) studies consider the accuracy of defect area is most important to estimate precisely the severity of the infraction. The defect area inside myocardial wall defined as the coldest spot in SPECT images where the blood does not per-fused through this spot. In our study we used fabricated phantom from polyethylene plastic consist of two chambers to simulate myocardial wall of the left ventricle in end-diastole and end-systole. 4 plastic defects in same thickness were inserted at 4 regions inside myocardial phantom. Clarity of the defect of acquired SPECT images is defined by thickness. Our work was carried out to measure the defect area of SPECT myocardial image with depending on the positioning of phantom on an imaging table. The results showed that clarity of defect of acquired SPECT depends on location of the defect related to myocardial thickness (end-diastole, end-systole), and on positioning of the phantom. The measured area of the defect in acquired SPECT image was compared with true defect area. The measured defect area in SPECT images in comparison with true defect area has varied results depending on the thickness of myocardial phantom(end-diastole , end-systole), and on the positioning of the phantom as follows: underestimated 6% and 4% at the center for end-diastole and end-systole respectively, and overestimation 2%, 5%, 9% and 1%, 3%, 6% at off-center  for end-diastole and end-systole respectivel that when the myocardial phantom positioned at 5 cm, 10 cm, and 15 cm on an imaging table.
References
Mendis S, Puska P, Norrving B. Global atlas on cardiovascular disease prevention and control: World Health Organization; 2011.
Greenwood JP, Maredia N, Younger JF, Brown JM, Nixon J, Everett CC, et al. Cardiovascular magnetic resonance and single-photon emission computed tomography for diagnosis of coronary heart disease (CE-MARC): a prospective trial. The Lancet. 2012;379(9814):453-60.
Berman DS, Kang X, Slomka PJ, Gerlach J, de Yang L, Hayes SW, et al. Underestimation of extent of ischemia by gated SPECT myocardial perfusion imaging in patients with left main coronary artery disease. Journal of Nuclear Cardiology. 2007;14(4):521-8.
Mahmarian JJ, Cerqueira MD, Iskandrian AE, Bateman TM, Thomas GS, Hendel RC, et al. Regadenoson induces comparable left ventricular perfusion defects as adenosine: a quantitative analysis from the ADVANCE MPI 2 trial. JACC: Cardiovascular Imaging. 2009;2(8):959-68.
Berti V, Sciagrà R, Acampa W, Ricci F, Cerisano G, Gallicchio R, et al. Relationship between infarct size and severity measured by gated SPECT and long-term left ventricular remodelling after acute myocardial infarction. European Journal of Nuclear Medicine and Molecular Imaging. 2011;38(6):1124-31.
Gibbons RJ. Assessment of infarct size and severity by quantitative myocardial SPECT: results from a multicenter study using a cardiac phantom. The Journal of Nuclear Medicine. 2000;41(8):1383.
Benoit T, Vivegnis D, Foulon J, Rigo P. Quantitative evaluation of myocardial single-photon emission tomographic imaging: application to the measurement of perfusion defect size and severity. European Journal of Nuclear Medicine and Molecular Imaging. 1996;23(12):1603-12.
Movahed A, Gnanasegaran G, Buscombe J, Hall M. Integrating Cardiology for Nuclear Medicine Physicians. Alemanha: Springer Verlag. 2009.
Cullom S, Case J, Bateman T. Attenuation correction in myocardial perfusion SPECT. Cardiac SPECT imaging 2nd ed Philadelphia: Lippincott Williams & Wilkins. 2001:89-102.
Zeng GL, Christian PE. Easy Method of Patient Positioning for Convergent-Beam Cardiac SPECT. Journal of nuclear medicine technology. 2007;35(3):131-4.
Sharma P, Patel CD, Karunanithi S, Maharjan S, Malhotra A. Comparative Accuracy of CT Attenuation-Corrected and Non–Attenuation-Corrected SPECT Myocardial Perfusion Imaging. Clinical nuclear medicine. 2012;37(4):332-8.
Lyra M, Sotiropoulos M, Lagopati N, Gavrilleli M, editors. Quantification of myocardial perfusion in 3D SPECT images-stress/rest volume differences. Imaging Systems and Techniques (IST), 2010 IEEE International Conference on; 2010: IEEE.
Hubbell JH, Seltzer SM. Tables of X-ray mass attenuation coefficients and mass energy-absorption coefficients 1 keV to 20 MeV for elements Z= 1 to 92 and 48 additional substances of dosimetric interest. National Inst. of Standards and Technology-PL, Gaithersburg, MD (United States). Ionizing Radiation Div., 1995.
Vallejo E. New radiotracers in cardiac imaging: Principles and applications taillefer and Tamaki. Journal of Nuclear Cardiology. 1999;6(5):549-50.
Knesaurek K, King MA, Glick SJ, Penney BC. Investigation of causes of geometric distortion in 180 degrees and 360 degrees angular sampling in SPECT. Journal of nuclear medicine: official publication, Society of Nuclear Medicine. 1989;30(10):1666-75.
Kirac S, Franc J, Liu Y-H. Validation of the Yale circumferential quantification method using (201) T1 and (99m) Tc: A phantom study. The Journal of Nuclear Medicine. 2000;41(8):14.
Downloads
Published
Issue
Section
License
- Papers must be submitted on the understanding that they have not been published elsewhere (except in the form of an abstract or as part of a published lecture, review, or thesis) and are not currently under consideration by another journal published by any other publisher.
- It is also the authors responsibility to ensure that the articles emanating from a particular source are submitted with the necessary approval.
- The authors warrant that the paper is original and that he/she is the author of the paper, except for material that is clearly identified as to its original source, with permission notices from the copyright owners where required.
- The authors ensure that all the references carefully and they are accurate in the text as well as in the list of references (and vice versa).
- Authors retain copyright and grant the journal right of first publication with the work simultaneously licensed under a Attribution-NonCommercial 4.0 International that allows others to share the work with an acknowledgement of the work's authorship and initial publication in this journal.
- Authors are able to enter into separate, additional contractual arrangements for the non-exclusive distribution of the journal's published version of the work (e.g., post it to an institutional repository or publish it in a book), with an acknowledgement of its initial publication in this journal.
- Authors are permitted and encouraged to post their work online (e.g., in institutional repositories or on their website) prior to and during the submission process, as it can lead to productive exchanges, as well as earlier and greater citation of published work (See The Effect of Open Access).
- The journal/publisher is not responsible for subsequent uses of the work. It is the author's responsibility to bring an infringement action if so desired by the author.